IRECTION FR091521, REVISION 1 VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
8-118 Section 8-5 - Electronic Cage Components - Replacement Procedures
PRELIMINARY
3.) Working from the rear of the Vivid S60N/Vivid S70N scanner, remove the four plastic air sleeve
securing screws - see Figure 8-144.
4.) Remove the CFE Plastic Air Sleeve.
8-5-13-5 Air Sleeve for CFE Installation Procedure
1) Position the replacement CFE Plastic Air Sleeve in its location in the cabinet cage.
2) Secure the part with the four securing screws previously removed.
3.) Close the cage door.
4.) Refit the covers previously remove:
Figure 8-144 Plastic Air Sleeve Securing Screws (rear of Scanner)
• Right Rear Cover Installation Procedure on page 8 - 22
• Left Rear Cover Installation Procedure on page 8 - 25
• Front Cover Installation Procedure on page 8 - 19
• Right Side Cover Installation Procedure on page 8 - 10
• Air Inlet (Left Side) Cover Installation Procedure on page 8 - 8