IRECTION FR091521, REVISION 1 VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-117
PRELIMINARY
8-5-13 Plastic Air Sleeve for CFE Replacement Procedure
8-5-13-1 Tools
Appropriate Phillips screwdriver.
8-5-13-2 Time Required
15 minutes
8-5-13-3 Preparation
Shut the Vivid™ S60/Vivid™ S70 ultrasound unit, as described in Power Shut Down on page 4 - 7.
8-5-13-4 Plastic Air Sleeve for CFE Removal Procedure
The Plastic Air Sleeve for CFE is located inside the cabinet cage - see Figure 8-143.
Figure 8-143 Location of Plastic Air Sleeve for CFE
1) Remove the following covers: left side, right side, front, left rear, right rear:
2.) Open the Cage door.
FRU Part #
Refer to Table 9-16 on page 9-15.
Air Inlet (Left Side) Cover Removal Procedure on page 8 - 7
Right Side Cover Removal Procedure on page 8 - 10
Front Cover Removal Procedure on page 8 - 16
Left Rear Cover Removal Procedure on page 8 - 23
Right Rear Cover Removal Procedure on page 8 - 20