235
────────────────────────────────────────────────────
11.4 Command Reference
────────────────────────────────────────────────────
■ Voltage range setting
Synta
VOLTage:RANGe data
dat
150, 300, 600 (numerical data in NR1 format)
Functio
Sets the voltage range. The units are volts (V).
Erro
・ In the following circumstances, a device dependent error is generated:
If this command is executed in the HOLD state .
If an attempt was made to execute this command while waiting for or
executing integrated measurement and demand measurement.
In the event of a system error.
・ data is other than NRf format data, an execution error is generated.
Exampl
Transmission
":VOLT:RANG 150"
The range is set to 150 V.
■ Voltage rang setting query
Synta
VOLTage:RANGe?
Functio
Returns the presently set voltage range as a numerical value in NR1
format.
Respons
synta
If headers are ON
If headers are OFF
":VOLTAGE:RANGE data"
"data"
Exampl
Transmission
Response
":VOLT:RANG?"
":VOLTAGE:RANGE 150"
":VOLT:RANG?"
"150"
OLTage:RANGe
OLTage:RANGe?