4.4 Judging Measurement Results
119
4
Chapter 4 LCR Function
The set reference value and upper and lower limit values are common to percentage mode
and percentage deviation mode.
Press .
7
Upper/Lower Limit Value Settings
Use the numeric keypad to enter the upper
limit value of the first parameter.
8
Settable range: -999.999% to 999.999%
Upper Limit Value Setting
Press to confirm the upper limit value.
When you do not want to set the upper and
lower limit values: Press .
9
Return to step 7, press , and use the numeric keypad to enter the lower limit value.
Settable range: -999.999% to 999.999%
10
Press to confirm the lower limit value.
11
Return to step 6, and set the upper and lower limit values of the third parameter in the same way.
12
Press to return to the BIN setting screen.
13
Press to close the setting screen.