8.5 Calculating Values (Scaling)
247
8
Chapter 8 Error Compensation
3
Select the compensation coefficient of the
parameter you want to change.
The parameters and compensation coefficient numbers
correspond as shown below.
Scaling Compensation Setting
Parameter 1
Parameter 2
Parameter 3
Parameter 4
4
Press .
Compensation Coefficient Setting
When you want to revert the settings to their
default values:
Press .
When you want to cancel the setting:
Press .
Use the numeric keypad to set compensation
coefficient A.
Settable range: -999.999 to 999.999
To return to the previous screen without making any change
to the set value, press the key when the screen is in
the state with nothing being displayed (the state after press-
ing the key).
If you make a mistake during input:
press to cancel the input and start again.
5
Compensation Coefficient Setting
Touch to accept compensation coeffi-
cient A.