EasyManua.ls Logo

Hioki IM3533 - Page 351

Hioki IM3533
400 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
14.2 Measurement Range and Accuracy
343
14
Chapter 14 Specifications
Accuracy table (IM3533 and IM3533-01)
The following value is added to the basic accuracy when performing temperature correction during DC resis-
tance measurement:
C Level coefficient
The coefficient corresponding to the setting for measurement level is obtained from the measurement level
coefficient table and then multiplied by the basic accuracy.
AC measurement
DC resistance measurement
Upper portion: Impedance Z (Unit: %) Lower portion: Phase angle (Unit: °)
Range DC
1 mHz to
99.999 Hz
100.00 Hz to
999.99 Hz
1.0000 kHz to
10.000 kHz
10.001 kHz to
100.00 kHz
100.01 kHz to
200.00 kHz
100 M
A= 1 B= 1 A= 6 B= 5 A= 3 B= 2 A= 3 B= 2 - - - -
A= 5 B= 3 A= 2 B= 2 A= 2 B= 2 - - - -
10 M
A= 0.5 B= 0.3 A= 0.8 B= 1 A= 0.5 B= 0.3 A= 0.5 B= 0.3 A= 3 B= 2 - -
A= 0.8 B= 0.5 A= 0.4 B= 0.2 A= 0.4 B= 0.2 A= 2 B= 2 - -
1 M
A= 0.2 B= 0.1 A= 0.4 B= 0.08 A= 0.3 B= 0.05 A= 0.3 B= 0.05 A= 0.7 B= 0.08 A= 1 B= 0.5
A= 0.3 B= 0.08 A= 0.2 B= 0.02 A= 0.2 B= 0.02 A= 1.5 B= 0.08 A= 3 B= 0.5
100 k
A= 0.1 B= 0.01 A= 0.3 B= 0.03 A= 0.2 B= 0.03 A= 0.15 B= 0.02 A= 0.25 B= 0.04 A= 0.4 B= 0.3
A= 0.3 B= 0.02 A= 0.1 B= 0.02 A= 0.1 B= 0.015 A= 0.4 B= 0.02 A= 1.2 B= 0.3
10 k
A= 0.1 B= 0.01 A= 0.3 B= 0.025 A= 0.2 B= 0.025 A= 0.05 B= 0.02 A= 0.2 B= 0.025 A= 0.3 B= 0.03
A= 0.3 B= 0.02 A= 0.1 B= 0.02 A= 0.03 B= 0.02 A= 0.4 B= 0.02 A= 0.6 B= 0.05
1 k
A= 0.1 B= 0.01 A= 0.3 B= 0.02 A= 0.2 B= 0.02 A= 0.15 B= 0.02 A= 0.2 B= 0.02 A= 0.3 B= 0.02
A= 0.2 B= 0.02 A= 0.1 B= 0.02 A= 0.08 B= 0.02 A= 0.4 B= 0.02 A= 0.6 B= 0.02
100
A= 0.1 B= 0.02 A= 0.4 B= 0.02 A= 0.3 B= 0.02 A= 0.15 B= 0.02 A= 0.2 B= 0.02 A= 0.3 B= 0.03
A= 0.2 B= 0.01 A= 0.15 B= 0.01 A= 0.1 B= 0.01 A= 0.4 B= 0.02 A= 0.6 B= 0.02
10
A= 0.2 B= 0.15 A= 0.5 B= 0.2 A= 0.4 B= 0.05 A= 0.3 B= 0.05 A= 0.3 B= 0.05 A= 0.4 B= 0.2
A=0.3 B=0.1 A=0.3 B=0.03 A=0.15 B=0.03 A=0.75 B=0.05 A=1.5 B=0.1
1
A= 0.3 B= 0.3 A= 2 B= 1 A= 0.6 B= 0.3 A= 0.4 B= 0.3 A= 0.4 B= 0.3 A= 1 B= 1
A= 1 B= 0.6 A= 0.5 B= 0.2 A= 0.25 B= 0.2 A= 1 B= 0.2 A= 2 B= 0.5
100 m
A= 3 B= 3 A= 10 B= 10 A= 3 B= 3 A= 3 B= 2 A= 2 B= 2 A= 4 B= 3
A= 6 B= 6 A= 2 B= 2 A= 2 B= 1.5 A= 2 B= 1.5 A= 3 B= 4
0.005 V to 0.999 V 1 V 1.001 V to 5 V
Level coefficient 1
V: Setting value (equivalent to when V mode) [V]
2 V
Level coefficient 1
-100
t0
t
1 +
t0
× (t + t - t
0
)
[%]
t0
: Basic temperature [°C]
t: Current temperature [
°C]
t: Temperature measurement accuracy
t0
:Temperature coefficient at t0 [1/ °C]
1+
0.2
V
1+
2
V

Table of Contents