14.3 About Measurement Times and Measurement Speed
352
• When the frequency is changed:
When the frequency is changed, the wait time is 1 ms.
However, a 3 ms wait time will be used if the measurement frequency change spans the two ranges of
(1 mHz to 39.999 Hz) and (40 Hz to 200 kHz).
• When switching levels
When the AC signal level is changed, the wait time is 1 ms.
•
When switching ranges
When the range is changed, the wait time is 1 ms.
Also, when low Z impedance high accuracy mode is turned ON/OFF, the wait time is 1 ms.
• When DC bias
When DC bias is changed to ON/OFF and when the DC bias level is changed, the wait time is 1 ms.
• When DC resistance measurement
When switching from AC measurement to DC resistance measurement, the wait time is 3 ms.
• When panel load
After all changes have been made, the maximum applicable wait time listed above will be used.
Wait time