14.3 About Measurement Times and Measurement Speed
352
When the frequency is changed:
When the frequency is changed, the wait time is 1 ms.
However, a 3 ms wait time will be used if the measurement frequency change spans the two ranges of
(1 mHz to 39.999 Hz) and (40 Hz to 200 kHz).
When switching levels
When the AC signal level is changed, the wait time is 1 ms.
When switching ranges
When the range is changed, the wait time is 1 ms.
Also, when low Z impedance high accuracy mode is turned ON/OFF, the wait time is 1 ms.
When DC bias
When DC bias is changed to ON/OFF and when the DC bias level is changed, the wait time is 1 ms.
When DC resistance measurement
When switching from AC measurement to DC resistance measurement, the wait time is 3 ms.
When panel load
After all changes have been made, the maximum applicable wait time listed above will be used.
Wait time