15.2 Troubleshooting
359
Chapter 15 Maintenance and Service
15
Perform a system reset.
This will return all settings to their factory defaults.
See (p. 147), (p.A18)
Performing a full reset will restore all of the settings to the factory default settings (p.A18).
Only perform a full reset in the following cases.
• When the normal reset screen cannot be displayed because of a problem with the instrument.
(After the full reset, perform a self check to confirm that there are no problems (p. 265).)
• When you have forgotten the passcode for the key lock.
When no apparent cause can be established
Full Reset Procedure
1
Turn ON the switch on the back panel.
2
While the opening screen is displayed, press
the top right of the screen continuously.
Connect the power cable.
3
If the instrument still does not operate normally after the full reset, it needs to be repaired.
Contact your dealer, or a Hioki representative if you are not sure where the instrument was purchased.
The full reset is complete when a beeping sound
is emitted.
4
Opening Screen
After the full reset, the panel calibration screen is
displayed automatically (p. 266).
Disconnect the measurement sample before performing a full reset.