4 CALibration Subsystem
116 U8480 Series Programming Guide
CALibration Command Subsystem
The CALibration command subsystem is used to zero and calibrate the U8480
Series.
The numeric suffix of the CALibration command (CALibration1) refers to Channel
A.
Zeroing and calibration of the U8480-Series is recommended:
–When a 5
°C change in temperature occurs
When connection to the U8480-Series is established
Every 24 hours
Prior to measuring low-level signals.
The following CALibration commands are overlapped commands:
CAL:ALL
CAL:AUTO
CAL:ZERO:AUTO
An overlapped command allows the U8480 Series to continue parsing and
executing subsequent commands while it is still executing.