2 MEASurement Commands
84 U8480 Series Programming Guide
READ[1][:SCALar][:POWer:AC]? 
[<expected_value>[,<resolution>[,<source list>]]]
This query sets the measurement function, aborts then initiates the measurement, 
calculates the measurement result, and places the result on the bus. The result is 
a power-based measurement and is expressed in the units defined by 
UNIT[1]:POWer.
Refer to “Auto-Averaging Settings” on page 374 to configure the correct 
parameters for expected power and resolution.
Syntax
INITiate[1]:CONTinuous must be set to OFF, otherwise error –213, “INIT 
ignored” occurs. If TRIGger[1]:SOURce is set to BUS, error –214, “Trigger 
deadlock” occurs.
:POW
:AC
:SCAL
?
Space
expected_value
DEF
resolution
,
DEF
source list
,
READ
1