EasyManua.ls Logo

Keysight U8480 Series - Trigger[:Sequence[1]]:Delay:auto <Boolean

Keysight U8480 Series
375 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
12 TRIGger Subsystem
322 U8480 Series Programming Guide
TRIGger[:SEQuence[1]]:DELay:AUTO <boolean>
This command is used to determine whether or not there is a settling-time delay
before a measurement is made.
When this command is set to:
ON, the U8480 Series inserts a settling-time delay before taking the requested
measurement and for subsequent measurements. This settling time allows the
internal digital filter to be updated with new values to produce valid, accurate
measurement results. The trigger with delay command allows settling time for
the internal amplifiers and filters.
In cases of large power changes, the delay may not be sufficient for complete
settling. Accurate readings can be assured by taking two successive
measurements for comparison.
OFF, no settling-time delay is inserted and the U8480 Series makes the
measurement immediately once a trigger is received.
Syntax
Example
TRIG:DEL:AUTO ON This command enables a delay on the U8480 Series.
:DEL
0|OFF
1|ON
Space
?
:AUTO
TRIG
:SEQ
1

Table of Contents