SENSe Subsystem 8
U8480 Series Programming Guide 179
[SENSe[1]:]AVERage:COUNt <numeric_value>
This command is used to enter a value for the filter length. If
[SENSe[1]:]AVERage:COUNt:AUTO is set to ON, then entering a value for the filter
length automatically sets it to OFF. Increasing the value of filter length increases
measurement accuracy but also increases the time taken to make a power
measurement.
Entering a value using this command automatically turns the
[SENSe[1]:]AVERage:STATe command to ON.
Syntax
NOTE
For most applications, automatic filter length selection
([SENSe[1]:]AVERage:COUNt:AUTO ON) is the best mode of operation. However,
manual filter length selection ([SENSe[1]:]AVERage:COUNt <numeric_value>) is
useful in applications requiring either high resolution or fast settling times,
where signal variations rather than measurement noise need filtering, or when
approximate results are needed quickly.
:COUN
Space
?
numeric_value
DEF
MIN
MAX
MAX
Space
SENS
1
AVER
:
MIN