8 SENSe Subsystem
218 U8480 Series Programming Guide
[SENSe[1]:]CORRection:CSET4:STATe <boolean>
This command is used to enable or disable S-Parameter correction.
When this is enabled, S-Parameter values from the selected S-Parameter table
will be used for correction.
Syntax
Reset condition
On preset (SYSTem:PRESet) and U8480 Series power-up,
[SENSe[1]:]CORRection:CSET4:STATe is set to ON.
Query
[SENSe[1]:]CORRection:CSET4:STATe?
The query returns the S-Parameter correction state.
Query example
Error message
– If you set this command to ON and no table has been selected using
[SENSe[1]:]CORRection:CSET4[:SELect], then error –221, “Settings
conflict” occurs and [SENSe[1]:]CORRection:CSET4:STATe remains OFF
SENS
1
:CSET4CORR
:
:STAT
0|OFF
1|ON
Space
?
CORR:CSET4:STAT? Queries the current state of the S-Parameter
correction.