EasyManua.ls Logo

Keysight U8480 Series - Table 1-6 Bit Definitions - Questionable Status Registers

Keysight U8480 Series
375 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
U8480 Series Remote Operation 1
U8480 Series Programming Guide 63
The following bits in these registers are used by the U8480 Series.
Table 1-6 Bit definitions - Questionable status registers
The condition bits are set and cleared under the following conditions:
Table 1-7 Bit change conditions for Questionable status register
Bit number Decimal weight Definition
0 to 2 - Not Used
3 8 POWer Summary
4 to 7 - Not Used
8 256 CALibration Summary
9 512 Power-On Self-Test
10 to 14 - Not Used
15 - Not Used (always 0)
Bit number Definition EVENts causing bit changes
3 POWer Summary This is a summary bit for the Questionable POWer Register
SET:
Error –230, “Data corrupt or stale”
Error –231, “Data questionable;Input Overload”
Error –231, “Data questionable;ZERO ERROR”
CLEARED:
When no errors are detected by the U8480 Series during a measurement
covering the causes given for it to set

Table of Contents