MEASurement Commands 2
U8480 Series Programming Guide 75
CONFigure[1] Command
The CONFigure command is used to set:
the expected measurement power level.
the measurement resolution.
The CONFigure command does not make the power measurement after setting
the configuration. Use READ?, or alternatively use INITiate followed by a FETCh?,
to make the measurement.
The CONFigure command also applies the following defaults to the
measurement(s) which are specified in the <source list> parameter:
Default settings Description
INITiate[1]:CONTinuous OFF Sets the U8480 Series to make one trigger cycle when
INITiate is sent
TRIGger[1]:SOURce IMMediate When TRIG:SOUR is set to BUS or HOLD, sets the U8480
Series to make the measurement immediately once a trigger
is received
TRIGger[1]:DELay:AUTO ON Enables automatic delay before making the measurement
[SENSe[1]:]AVERage:COUNt:AUTO ON Enables automatic filter length selection
[SENSe[1]:]AVERage:STATe ON Enables averaging