12 TRIGger Subsystem
306 U8480 Series Programming Guide
INITiate[1]:CONTinuous:SEQuence[1] <boolean>
This command sets the U8480 Series for either a single trigger cycle or
continuous trigger cycles. A trigger cycle means that the U8480 Series exits the
wait-for-trigger state and starts a measurement.
If INITiate[1]:CONTinuous:SEQuence[1] <boolean> is set to:
– OFF, the trigger system remains in the idle state until it is set to ON or
INITiate[1]:IMMediate is received. Once this trigger cycle is complete, the
trigger system returns to the idle state.
– ON, the trigger system is initiated and exits the idle state. On completion of
each trigger cycle, the trigger system immediately commences another trigger
cycle without entering the idle state.
Syntax
Example
Reset condition
On reset (*RST), this command is disabled.
On preset (SYSTem:PRESet) and U8480 Series power-up, this command is
enabled.
This command performs the same functions as INITiate[1]:CONTinuous
<boolean>.
INIT:CONT:SEQ ON This command places the U8480 Series in a
wait-for-trigger state.
INIT
:CONT
0|OFF
1|ON
Space
?
:SEQ
1
1