SENSe Subsystem 8
U8480 Series Programming Guide 217
[SENSe[1]:]CORRection:CSET3:[SELect] <“string”>
This command is used to select the Gamma table to be used in the Table-Based
Gamma correction.
Syntax
Query
[SENSe[1]:]CORRection:CSET3:[SELect]?
The query returns the currently selected Gamma table used in the Table-Based
Gamma correction.
If no table is selected, the query returns an empty string.
Query example
Error message
If the string is not valid, error –224, “Illegal parameter value” occurs.
If a table does not exist, error –256, “File name not found” occurs.
When a gamma table is selected, the power sensor verifies that the number of
magnitude-phase offset pairs defined is equal to the number of frequency
points defined. If they do not match, error –226, “Lists not the same length”
occurs.
SENS
1
:CSET3
CORR
:
?
:SEL
“string”
Space
CORR:CSET3:SEL? Queries the current table used in the Table-Based
Gamma correction.