1 U8480 Series Remote Operation
28 U8480 Series Programming Guide
Making Measurements
The MEASure? query and CONFigure command provide a straightforward method 
to program the U8480 Series for measurements. You can select the measurement 
expected power level and resolution in one command. The U8480 Series 
automatically presets other measurement parameters to default values as shown 
in Table 1-1 below. 
Table 1-1 MEASure? and CONFigure preset states
An alternative method to program the U8480 Series is to use the lower-level 
commands. The advantage of using the lower-level commands over the 
MEASure? query and CONFigure command is that they give you more precise 
control of the U8480 Series. As shown in Table 1-1, the CONFigure command 
presets various states in the U8480 Series. It may be likely that you do not want to 
preset these states.
Using MEASure?
The simplest way to program the U8480 Series for measurements is by using the 
MEASure? query. However, this query does not offer much flexibility. When you 
execute the query, the U8480 Series selects the best settings for the requested 
configuration and immediately performs the measurement. You cannot change any 
setting (other than the expected power value and resolution) before the 
measurement is taken. This means you cannot finetune the measurement; for 
Command MEASure? and CONFigure settings
Trigger source
(TRIGger[1]:SOURce)
Immediate
Filter
([SENSe[1]:]AVERage:COUNt:AUTO)
On
Filter state
([SENSe[1]:]AVERage[:STATe])
On
Trigger cycle
(INITiate[1]:CONTinuous)
Off
Trigger delay
(TRIGger[1]:DELay:AUTO)
On