2 MEASurement Commands
74 U8480 Series Programming Guide
CONFigure[1]?
This query returns the present configuration of the measurement.
Syntax
The returned string depends on the setting of the CALCulate:MATH commands.
The configuration is returned as a quoted string in the following format:
<function> <expected_value>,<resolution>,<source list>”
<expected_value> returns the expected value sent by the last CONFigure
command or +20 dBm by default.
Example
Reset condition
On reset:
The command function is set to :POWer:AC.
The expected power level is set to +20 dBm.
The resolution is set to 3.
The source list on the U8480 Series is set to (@1).
?
CONF
1
CONF? Queries the measurement configuration.