TRIGger Subsystem 12
U8480 Series Programming Guide 311
TRIGger[1]:DELay:AUTO <boolean>
This command is used to determine whether or not there is a settling-time delay
before a measurement is made.
When this command is set to:
– ON, the U8480 Series inserts a settling-time delay before taking the requested
measurement. This settling time allows the internal digital filter to be updated
with new values to produce valid and accurate measurement results. The
trigger with delay command allows settling time for the internal amplifiers and
filters.
– In cases of large power changes, the delay may not be sufficient for a complete
settling. Accurate readings can be assured by taking two successive
measurements for comparison.
– OFF, the U8480 Series makes the measurement immediately after a trigger is
received.
Syntax
Example
TRIG:DEL:AUTO ON This command enables a delay on the U8480 Series.
TRIG
:AUTO
:DEL 0|OFF
1|ON
Space
?
1