MEASurement Commands 2
U8480 Series Programming Guide 91
MEASure[1][:SCALar][:POWer:AC]:MUNC?
[<expected_value>[,<resolution>[,<source list>]]]
This command aborts any measurement in progress, configures the U8480 Series,
calculates the measurement result and the corresponding measurement
uncertainty, and places the result on the bus. The result is a power-based
measurement and is expressed in the units defined by UNIT[1]:POWer.
When “UNIT:POW W” is set, the parameters returned will be:
–measured power in Watts
+measurement uncertainty value in %
–measurement uncertainty value in %
–WhenUNIT:POW DBM” is set, the parameters returned will be:
–measured power in dBm
+measurement uncertainty value in dB
–measurement uncertainty value in dB.
Syntax
NOTE
This query is only supported for TRIG:COUNT of 1. It is not supported for power
and frequency sweep modes.
:POW
:AC
?
:SCAL
Space
expected_value
DEF
resolution
,
DEF
source list
,
MEAS
1
:MUNC