2 MEASurement Commands
92 U8480 Series Programming Guide
Parameters
Refer to “Optional parameters” on page 73 for additional details on the
parameters in this query.
Error messages
If this query is sent, and TRIG:COUNT > 1, error –221, “Settings conflict”
occurs.
Item Description/Default Range of values
expected_value
(for the expected power
level)
A numeric value for the expected power level. The units of
measurement are dBm and W. The default units are defined
by UNIT:POWer.
sensor-dependent
DEF
[a]
resolution A numeric value for the resolution. If unspecified, the current
resolution setting is used.
1 to 4
[b]
1.0, 0.1, 0.01, 0.001
DEF
[a]
source list The measurement channel which the command is
implemented on. The U8480 Series supports only one
channel. Therefore, the only valid value is (@1).
(@1)
[a] The mnemonic DEF means DEFault. This is not equivalent to the DEFault parameter used in the command subsystems. The parameters must
be entered in the specified order. If parameters are omitted, they default from the right. The parameter DEFault is used as a placeholder.
Specifying DEF leaves the parameter value unchanged.
[b] When the measurement result is linear, this parameter represents the number of significant digits. When the measurement result is
logarithmic, 1 to 4 represents 1, 0.1, 0.01, and 0.001 respectively.