3 CALCulate Subsystem
104 U8480 Series Programming Guide
CALCulate[1]:LIMit:LOWer[:DATA] <numeric_value>
This command enters a value for the lower test limit for the measurement used in
the CALCulate[1]:LIMit:FAIL? test. The units used are dependent on the
current setting of UNIT:POWer. When the measured value is less than the value
specified in CALCulate[1]:LIMit:LOWer[:DATA], CALCulate[1]:LIMit:FAIL?
reports a fail. When the measured value is greater than or equal to the limit, a fail
is not reported.
Syntax
Parameters
Item Description/Default Range of values
numeric_value A numeric value for the lower test limit:
– DEF: The default is –90.00 dBm
– MIN: –150 dBm
– MAX: +230 dBm
–150 to +230 dBm
DEF
MIN
MAX
Space
?
numeric_value
DEF
MIN
MAX
MIN
MAX
Space
:LIM
:LOW
:DATA
CALC
1