TRIGger Subsystem 12
U8480 Series Programming Guide 299
ABORt[1]
This command removes the U8480 Series from the wait-for-trigger state and
places it in the idle state. It does not affect any other settings of the trigger
system. When the INITiate command is sent, the trigger system responds as it
did before ABORt[1] was executed.
If INITiate[1]:CONTinuous is ON, then after ABORt[1], the measurement
immediately goes into the wait-for-trigger state.
Syntax
Example
ABOR This command places the U8480 Series in
the idle state.