TRIGger Subsystem 12
U8480 Series Programming Guide 325
TRIGger[:SEQuence[1]]:SOURce BUS|EXTernal|HOLD|IMMediate
This command configures the trigger system to respond to the specified source.
This command only selects the trigger source. Use the INITiate command to
place the U8480 Series in the wait-for-trigger state.
Syntax
Parameters
NOTE
This command has the same purpose as TRIGger[1]:SOURce
BUS|EXTernal|HOLD|IMMediate.
Item Description/Default Range of values
source
Available trigger sources:
BUS: The trigger source is a *TRG common command or the
TRIGger[1][:IMMediate] SCPI command.
EXTernal: The trigger source is the trigger input in the U8480 Series
HOLD: Triggering is suspended. The only way to trigger the U8480 Series is to use
TRIGger[1][:IMMediate].
IMMediate: The trigger system is always true. If INITiate[1]:CONTinuous is
ON, the U8480 Series is continually triggering free (free run mode). If an
INITiate[1][:IMMediate] command is sent, a measurement is triggered and
then the U8480 Series returns to the idle state.
BUS
EXTernal
HOLD
IMMediate
:SOUR
TRIG
:SEQ
1
Space
?
BUS
EXT
HOLD
IMM