8 SENSe Subsystem
208 U8480 Series Programming Guide
[SENSe[1]:]CORRection:SGAMma:STATe <boolean>
This command is used to enable or disable Single Point Gamma correction.
Values from [SENSe[1]:]CORRection:SGAMma:MAGNitude and
[SENSe[1]:]CORRection:SGAMma:PHASe apply across all frequency values and
are used for correction when this is enabled.
Syntax
Reset condition
On preset (SYSTem:PRESet) and U8480 Series power-up,
[SENSe[1]:]CORRection:SGAMma:STATe is set to OFF.
Query
[SENSe[1]:]CORRection:SGAMma:STATe?
The query returns the Single Point Gamma correction state.
SENS
1
CORR
:
:STAT
:SGAM
0|OFF
1|ON
?
Space