SENSe Subsystem 8
U8480 Series Programming Guide 213
[SENSe[1]:]MUNC:SGAMma:TYPE? SINGle|TABLe|SPARam
The command selects the type of source Gamma that will be used in the
calculation of the Device-Under-Test (DUT) mismatch for the Measurement
Uncertainty feature. There are three types of source gamma that can be selected;
Single Point Gamma, Table-Based Gamma, and the S-Parameter Table.
The default source gamma is Single Point Gamma.
The value of [SENSe[1]:]MUNC:SGAM:TYPE will automatically change to reflect
the setting of [SENSe[1]:]CORRection:SGAMma:STATe and
[SENSe[1]:]CORRection:CSET3:STATe, unless its current selection is SPARam
If [SENSe[1]:]CORRection:CSET4:STATe is set to ON,
[SENSe[1]:]MUNC:SGAM:TYPE will change to SPARam.
Syntax
Reset condition
On preset (SYSTem:PRESet) and U8480 Series power-up,
[SENSe[1]:]MUNC:SGAM:TYPE is set to SINGle.
If there is a 2-port device connected to the sensor, select SPARam as the source
gamma.
SENS
1
MUNC
:
:SGAM
SING
TABL
SPAR
:TYPE
?