SENSe Subsystem 8
U8480 Series Programming Guide 233
[SENSe[1]:]MRATe <character_data>
This command sets the measurement rate on the U8480 Series.
The U8480 Series can operate in three different measurement rates as shown in
the following table.
[a] Minimum acquisition time is the minimum time for the U8480 Series to complete a measurement cycle. The time includes integration time
and settling time.
[b] In the Fast mode, the minimum acquisition time is the same as integration time. Integration time is defined as the period during which the
analog-to-digital (A/D) converter of the U8480 Series samples the input signal for a measurement.
When the measurement rate is set to Fast, the following couplings occur:
[a] This change only occurs on the measurement specified in the [SENSe[1]:]MRATe command. When the measurement rate
is changed from Fast to either Double or Normal, the settings that were in place when Fast was entered are restored.
Syntax
Measurement rate Measurement speed
Minimum acquisition time
[a]
Normal 20 readings/s 51.2 ms/reading
Double 40 readings/s 24.576 ms/reading
Fast
[b]
400 readings/s 2.048 ms/reading
Command Status
[SENSe[1]:]AVERage:STATe OFF
[a]
[SENSe[1]:]CORRection:DCYCle:STATe OFF
[a]
[SENSe[1]:]CORRection:GAIN2:STATe OFF
[a]
CALCulate[1]:MATH:EXPRession “(SENS1)”
SENS
1
:
Space
?
character_data
MRAT