8 SENSe Subsystem
234 U8480 Series Programming Guide
Parameters
Example
Reset condition
On reset, the measurement rate is set to NORMal.
Item Description/Default Range of values
character_data A numeric value for the measurement :
– NORMal: 20 readings/s
– DOUBle: 40 readings/s
– FAST: up to 400 readings/s
[a]
– The default is NORMal.
NORMal
[a] [d]
DOUBle
[b] [d]
FAST
[c] [d]
[a] The measurement is taken with SENS:AVER:STAT set to OFF.
[b] When NORMal or DOUBle is set, TRIGger[:SEQuence[1]]:COUNt is set automatically to 1.
[c] To reduce the sensor-dependent delay time, use the measurement buffer by setting TRIGger[:SEQuence[1]]:COUNt
higher than 1.
[d] Calibration of the U8480 Series is recommended after measurement rate is changed.
MRAT DOUB This command sets the measurement rate to 40 readings/s.