10 STATus Subsystem
268 U8480 Series Programming Guide
STATus:OPERation
The operation status register set contains conditions which are part of the U8480 
Series operation as a whole. 
The following bits in these registers are used by the U8480 Series:
Syntax
Bit number Decimal weight Definition
0 1 CALibrating summary
1 to 3 - Not used
4 16 MEASuring summary
5 32 Waiting for TRIGger summary
6 to 9 - Not used
10 1024 SENSe summary
11 2048 Lower limit fail summary
12 4096 Upper limit fail summary
13 to 15 - Not used (bit 15 is always 0)