EasyManua.ls Logo

Keysight U8480 Series - Status:questionable:calibration[:Summary]

Keysight U8480 Series
375 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
10 STATus Subsystem
278 U8480 Series Programming Guide
STATus:QUEStionable:CALibration[:SUMMary]
The questionable calibration summary register set contains bits which give an
indication of the quality of the data produced by the U8480 Series due to its
calibration status.
The following bit in these registers is used by the U8480 Series:
Bit 1 is set when:
CALibration[1]:ZERO:AUTO ONCE, CALibration[1]:AUTO ONCE,
CALibration[1][:ALL], or CALibration[1][:ALL]? is executed
error –231, “Data questionable;ZERO ERROR” or –231, “Data
questionable;CAL ERROR” occurs
This bit is cleared when any of the commands listed above succeed and no
errors are placed on the error queue.
Syntax
Bit number Decimal weight Definition
0-Not used
1 2 Summary of CALibration
2 to 15 - Not used (bit 15 is always 0)
STAT
:QUES
:CAL
:SUMM

Table of Contents

Related product manuals