STATus Subsystem 10
U8480 Series Programming Guide 269
STATus:OPERation:CALibrating[:SUMMary]
The operation status calibrating summary register set contains information on the
calibrating status of the U8480 Series.
The following bit in these registers is used by the U8480 Series:
Bit 1 is set at the beginning of zeroing (CALibration:ZERO:AUTO ONCE) or
calibration (CALibration:AUTO ONCE). Also for the compound command/query
CALibration[:ALL]?, this bit is set at the beginning of the calibration sequence.
This bit is cleared at the end of zeroing or calibration.
Syntax
Bit number Decimal weight Definition
0-Not used
1 2 CALibrating status
2 to 15 - Not used (bit 15 is always 0)
STAT
:OPER
:CAL
:SUMM