TRIGger Subsystem 12
U8480 Series Programming Guide 319
TRIGger[:SEQuence]:SLOPe <character_data>
This command specifies whether a trigger event is recognized on the rising or
falling edge of a signal.
Syntax
Parameters
Reset condition
On reset, the value is set to POSitive.
Query
TRIGger[:SEQuence]:SLOPe?
The query returns the current value of <character_data>.
Query example
Item Description/Default Range of values
character_data
How a trigger event is recognized:
– POSitive: A trigger event is recognized on the rising edge of a
signal
– NEGative: A trigger event is recognized on the falling edge of a
signal
POSitive
NEGative
:SLOP
Space
?
character_data
TRIG
:SEQ
TRIG:SLOP?
Queries the current value of <character_data>
for the U8480 Series.