3 CALCulate Subsystem
106 U8480 Series Programming Guide
CALCulate[1]:LIMit:UPPer[:DATA] <numeric_value>
This command enters a value for the upper test limit for the measurement used in
the CALCulate[1]:LIMit:FAIL? test. The units used are dependent on the
current setting of UNIT:POWer. When the measured power is greater than the
value specified in CALCulate[1]:LIMit:UPPer[:DATA],
CALCulate[1]:LIMit:FAIL? reports a fail. When the measured level is less than
or equal to the limit, a fail is not reported.
Syntax
Parameters
Item Description/Default Range of values
numeric_value A numeric value for the upper test limit:
DEF: The default is +90.00 dBm
MIN: –150 dBm
MAX: +230 dBm
–150 to +230 dBm
DEF
MIN
MAX
Space
?