MEMory Subsystem 6
U8480 Series Programming Guide 163
MEMory:TABLe:SELect <“character_data”>
This command is used to activate a frequency-dependent offset table, a Gamma
table, or an S-Parameter table. A table must be activated before any operation
can be performed on it.
Syntax
Parameters
Example
Query
MEMory:TABLe:SELect?
The query returns the name of the currently selected table.
Item Description/Default Range of values
character_data
Details the new table name. A maximum of 12 characters
can be used.
A to Z (uppercase)
a to z (lowercase)
0 to 9_ (underscore)
MEM:TABL:SEL “Sensor1” This command selects a
frequency-dependent offset table named
“Sensor1”.
MEM
:TABL
:SEL
?
Space
“character_data”