Setting Offsets
Channel offsets
The U8480 Series can be configured to compensate for signal loss or gain in your
test setup (for example, to compensate for the loss of a 10 dB attenuator). You use
the SENSe command subsystem to configure the U8480 Series. Gain and loss
correction are a coupled system. If you enter an offset value, the state is
automatically enabled. However, it can be enabled and disabled using the
[SENSe[1]:]CORRection:GAIN2:STATe <boolean> command.
NOTE
To enter a LOSS value, you can enter a negative value in the command:
[SENSe[1]:]CORRection:GAIN2[:INPut][:MAGNitude] <numeric_value>