1 U8480 Series Remote Operation
44 U8480 Series Programming Guide
Filter length
You specify the filter length using the following command:
[SENSe[1]:]AVERage:COUNt <numeric_value>
The range of values for the filter length is 1 to 1024. Specifying this command
disables automatic filter length selection. Increasing the value of the filter length
reduces measurement noise. However, the time to take the measurement is
increased.
Table 1-2 Settling time for normal speed, ×2 speed, and fast speed
Number of averages
Settling time
[a]
(s)
(Normal speed)
[a] Manual filter, 10 dB decreasing power step
Settling time
[a]
(s)
(´2 speed)
Settling time
[a]
(s)
(Fast speed)
1 0.15 0.14 0.003
2 0.23 0.16 0.005
4 0.33 0.23 0.009
8 0.53 0.33 0.018
16 0.90 0.51 0.036
32 1.68 0.91 0.069
64 3.24 1.70 0.134
128 6.44 3.28 0.265
256 12.7 6.45 0.528
512 25.3 12.7 1.05
1024 50.5 25.3 2.10