CALibration Subsystem 4
U8480 Series Programming Guide 123
CALibration[1]:TYPE EXTernal|INTernal
This command sets the U8480 Series to the external or internal calibration mode.
External calibration requires a 50 MHz 1 mW power reference, while internal
calibration utilizes the internal reference circuit to perform calibration and does
not require the 50 MHz 1 mW power reference.
Upon power up, the U8480 Series defaults to the internal calibration mode.
Syntax
Example
Reset condition
On reset, the calibration mode is set to internal.
NOTE
Setting CALibration[1]:TYPE to EXTernal will automatically set
CALibration[1]:AUTO to OFF.
CAL:TYPE EXT This command sets the external calibration mode.