8 SENSe Subsystem
180 U8480 Series Programming Guide
Parameters
Example
Reset condition
On reset, the filter length is set to four.
Query
[SENSe[1]:]AVERage:COUNt? [MIN|MAX]
The query returns the current setting of the filter length or the values associated
with MIN and MAX. The format of the response is <NR1>.
Query example
Error message
If a filter length value entered is outside the allowable range of values, error
–222, “Data out of range” occurs.
If a filter length value is entered using [SENSe[1]:]AVERage:COUNt while
[SENSe[1]:]MRATe is set to FAST and TRIG:COUNt > 1, error –221, “Settings
conflict” occurs. However, the filter length value is set but the
[SENSe[1]:]AVERage:STATe command is not automatically set to ON.
Item Description/Default Range of values
numeric_value
A numeric value defining the filter length:
DEF: The default value is 4
MIN: 1
MAX: 1024
1 to 1024
DEF
MIN
MAX
AVER:COUN 400 This command sets a filter length of 400.
AVER:COUN? Queries the filter length.