MEMory Subsystem 6
U8480 Series Programming Guide 157
Example
Query
MEMory:TABLe:FREQuency?
The query returns a list of frequency points for the table currently selected. The
frequencies are returned in Hz.
Query example
Error messages
If more than 80 frequencies for a frequency-dependant offset table, or more
than 1024 frequencies a Gamma or S-Parameter table are in the list, error
–108, “Parameter not allowed” occurs.
If the frequencies are not entered in the ascending order, error
–220, “Parameter error; Frequency list must be in ascending order” occurs.
If a table has not been specified using the MEMory:TABLe:SELect command,
the data cannot be entered into the table and error
–221, “Settings conflict” occurs.
If a frequency is set outside of the allowed frequency range, error
–222, “Data out of range” occurs.
MEM:TABL:FREQ 200MHz,600MHz This command enters frequencies of
200 MHz and 600 MHz into the currently
selected table.
MEM:TABL:FREQ? Queries the frequency points in the currently
selected table.