6 MEMory Subsystem
160 U8480 Series Programming Guide
Example
Query
MEMory:TABLe:GAIN[:MAGNitude]?
The query returns a list of offset points for the currently selected table.
Query example
Error messages
If more than 80 offsets for the frequency-dependent offset tables are in the list,
error –108, “Parameter not allowed” occurs.
If a table is not specified using the MEMory:TABLe:SELect command, the data
cannot be entered and error –221, “Settings conflict” occurs.
If any of the offset factors are outside of the allowed range, error
–222, “Data out of range” occurs.
MEM:TABL:SEL “Sensor_1”
MEM:TABL:GAIN 97,99.5,97.4
This command enters a reference offset
factor of 97%, 99.5%, and 97.4% into the
frequency-dependent offset table.
MEM:TABL:GAIN? Queries the offset in the current table.