8 SENSe Subsystem
188 U8480 Series Programming Guide
Query
[SENSe[1]:]BUFFer:COUNt? [MIN|MAX]
This query returns the average trigger measurement buffer size or the values
associated with MIN and MAX.
Query example
Error messages
–If TRIGger:SOURce is not set to EXT, error –221, “Settings conflict” occurs.
–If the [SENSe[1]:]BUFFer:COUNt parameter is set <1, error –222, “Data out
of range” occurs.
–If the [SENSe[1]:]BUFFer:COUNt parameter is set >250, error –222, “Data
out of range” occurs.
– If the frequency sweep step is non-zero, error –221, “Settings conflict” occurs.
BUFF:COUN? This query returns the average trigger measurement buffer
size.