SENSe Subsystem 8
U8480 Series Programming Guide 191
Query
[SENSe[1]:]CORRection:CSET2[:SELect]?
The query returns the name of the selected table as a quoted string. If no table is 
selected, an empty string is returned.
Query example
Error messages
–If <“string”> is invalid, error –224, “Illegal parameter value” occurs.
– If a table called <“string”> does not exist, error –256, “File name not found” 
occurs.
– When a frequency-dependent offset table is selected, the U8480 Series 
verifies that the number of offset points defined is equal to the number of 
frequency points defined. If this is not the case, error –226, “Lists not the same 
length” occurs.
CORR:CSET2? Queries the frequency-dependent offset table currently used.