15 Programming Examples
372 U8480 Series Programming Guide
-> 
MEM:TABL:SPAR S21,0.3,90 // Inputs the S21 magnitude-phase pair 
values; 0.3 for magnitude, and 90
o
 for 
phase
-> 
MEM:TABL:SPAR S22,0.7,–135 // Inputs the S22 magnitude-phase pair 
values; 0.7 for magnitude, and –135
o
 
for phase
-> 
SENS:CORR:CSET4:STAT OFF // Turns off S-Parameter Correction in 
order to include the mismatch between 
the DUT and the sensor in the MU 
calculation
-> 
SENS:CORR:CSET4:SEL “SParam1” // Selects “SParam1” as the table to be 
used to obtain the DUT Gamma
-> 
SENS:MUNC:SGAM:TYPE SPAR // Selects S-Parameter as the DUT 
Gamma
-> 
FREQ 50MHz // Sets the measurement frequency to 50 
MHz
-> 
SENS:MUNC:STAT ON // Turns on the RTMU function
-> 
FETCH:MUNC? // Queries the measured power and 
calculated MU