1 U8480 Series Remote Operation
42 U8480 Series Programming Guide
READ? example
ABOR
CONF DEF,2,(@1)
CORR:CSET2:SEL “Offset1”
CORR:CSET2:STAT ON
FREQ 50MHZ
READ?
NOTE
If the measurement frequency does not correspond directly to a frequency in the
frequency- dependent offset table, the U8480 Series calculates the offset using
linear interpolation.
If you enter a frequency outside the frequency range defined in the
frequency-dependent offset table, then the U8480 Series uses the highest or
lowest frequency point in the table to set the offset.
To find out the value of the offset being used by the U8480 Series to make a
measurement, use the following query:
[SENSe[1]:]CORRection:FDOFfset|GAIN4[:INPut][:MAGNitude]?
The response may be an interpolated value.