CALCulate Subsystem 3
U8480 Series Programming Guide 99
Reset condition
On reset, CALCulate[1]:LIMit:CLEar:AUTO is set to ON.
Query
CALCulate[1]:LIMit:CLEar:AUTO?
The query enters a 1 or 0 into the output buffer indicating whether limit failures
are cleared automatically when a new measurement is initiated.
1 is entered into the output buffer when limit failures are cleared automatically
when a new measurement is initiated.
0 is entered into the output buffer when limit failures are not cleared
automatically when a new measurement is initiated.
In the case where limit failures are cleared once, when a query occurs, 1 is entered
into the output buffer if no measurement is initiated. If a measurement is initiated,
then 0 is entered.
Query example
CALC:LIM:CLE:AUTO? Queries when the fail counter is cleared.