Receiver Mode
R&S
®
ESR
195User Manual 1175.7068.02 ─ 12
Remote command:
not supported
Sort by Delta Limit
Sort the entries of the peak list according to the delta limit results.
The delta limit is the distance of a peak to a limit line, if one has been assigned.
Remote command:
not supported
Symbols (On Off)
Turns the labels on the peak position in the diagram on and off.
The peak labels have a different color and shape depending on the trace they are on.
Trace 1, for example, has red crosses as the peak label. By default they are on.
Remote command:
DISPlay[:WINDow<n>]:TRACe<t>:SYMBol on page 682
Peak List Export
Opens a dialog box to export and save the contents of the peak list in ASCII format to
a *.dat file.
The file consists of a header and the results of the scan or the final measurement.
The header is a list of general instrument settings and characteristics. It consists of
three columns, each column separated by a semicolon: <parame-
ter>;<value>;<unit>.
The results are split into several data sections, one for each active trace. The data
section begins with the entry Trace <x> [Final]:, followed by the trace char-
actersistics and the peak list data itself.
For a description of the data see Chapter 6.3.1.6, "ASCII File Export Format",
on page 399 .
By default, decimal places are separated by a point in the exported list. If required, you
can use a comma instead of a point as the decimal separator.
Remote command:
Export the peak list of the scan results:
MMEMory:STORe:PEAKlist on page 682
Export the peak list of the final measurement:
MMEMory:STORe:FINal on page 682
Select the decimal separator:
FORMat:DEXPort:DSEParator on page 850
5.3.7.5 Trace / Final Meas
The Trace / Final Meas tab in the Test Automation dialog box contains functionality to
configure the traces for the scan and the final measurement.
Final Measurement Time.............................................................................................196
Interactive Mode..........................................................................................................196
Trace 1 to 6................................................................................................................. 196
Common Measurement Settings

Table of Contents

Related product manuals