Remote Control
R&S
®
ESR
802User Manual 1175.7068.02 ─ 12
Manual operation: See "Time Domain Power (zero span)" on page 224
See "Time Domain Power" on page 296
Retrieving Measurement Results
The following commands analyze and retrieve measurement results for time domain
power measurements.
Commands useful to retrieve results described elsewhere:
●
TRACe<n>:DATA on page 726
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:MEAN:AVERage:RESult?....................... 802
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:MEAN:PHOLd:RESult?.......................... 803
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:MEAN:RESult?......................................803
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:MEAN[:STATe].......................................804
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:PPEak:AVERage:RESult?...................... 804
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:PPEak:PHOLd:RESult?..........................805
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:PPEak:RESult?..................................... 805
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:PPEak[:STATe]...................................... 806
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:RMS:AVERage:RESult?......................... 806
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:RMS:PHOLd:RESult?.............................807
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:RMS:RESult?........................................ 808
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:RMS[:STATe].........................................808
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:SDEViation:AVERage:RESult?................809
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:SDEViation:PHOLd:RESult?................... 809
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:SDEViation:RESult?...............................810
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:SDEViation[:STATe]................................810
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:MEAN:AVERage:RESult?
This command queries the average mean time domain power. The query is only possi-
ble if averaging has been activated previously using CALCulate<n>:MARKer<m>:
FUNCtion:SUMMary:AVERage.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweeps.
Suffix:
<n>
.
Selects the measurement window.
<m> Selects the marker.
Return values:
<MeanPower> Mean power of the signal during the measurement time.
Remote Commands in Spectrum Analyzer Mode