Signal Count
CALCulate<n>:MARKer<m>:COUNt................................................................................870
CALCulate<n>:MARKer<m>:COUNt:FREQuency?............................................................870
CALCulate<n>:MARKer<m>:COUNt:RESolution...............................................................871
CALCulate<n>:MARKer<m>:COUNt <State>
This command turns the frequency counter at the marker position for marker 1 on and
off.
The count result is queried with CALCulate<n>:MARKer<m>:COUNt:FREQuency?.
The frequency counter works for one marker only. If you perform a frequency count
with another marker, the R&S ESR deactivates the frequency count of the first marker.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweeps.
Suffix:
<n>
.
Selects the measurement window.
<m> irrelevant
Parameters:
<State> ON | OFF
*RST: OFF
Example:
INIT:CONT OFF
Switches to single sweep mode.
CALC:MARK ON
Switches on marker 1.
CALC:MARK:COUN ON
Switches on the frequency counter for marker 1.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK:COUN:FREQ?
Outputs the measured value.
Manual operation: See "Signal Count" on page 417
CALCulate<n>:MARKer<m>:COUNt:FREQuency?
This command queries the result of the frequency counter for marker 1.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweeps.
Suffix:
<n>
.
Selects the measurement window.
<m> irrelevant
Remote Commands in Spectrum Analyzer Mode

Table of Contents