EasyManua.ls Logo

R&S ESR Series

R&S ESR Series
1126 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control
R&S
®
ESR
949User Manual 1175.7068.02 ─ 12
Usage: Event
SCPI confirmed
Manual operation: See "Save File / Recall File" on page 531
MMEMory:STORe<n>:STATe:NEXT
This command stores the current device settings in a *.dfl file. The file name used in
the last saving process is automatically counted up to the next unused name.
Suffix:
<n>
.
irrelevant
Example:
MMEM:STOR:STAT 1,'Save'
Saves the current device settings in the file Save.dfl.
MMEM:STOR:STAT:NEXT
Saves the current device settings in the file Save_001.dfl
MMEM:STOR:STAT:NEXT
Saves the current device settings in the file Save_002.dfl
Usage: Event
Manual operation: See "Save File / Recall File" on page 531
11.11 Using Transducers
Before making any changes to a transducer factor or set, you have to select one by
name with [SENSe:]CORRection:TRANsducer:SELect.
Compared to manual configuration of transducers, any changes made to a transducer
factor or set via remote control are saved after the corresponding command has been
sent.
Designing a transducer factor
//Select a transducer factor
CORR:TRAN:SEL 'Transducer1'
//Define a comment for the transducer factor
CORR:TRAN:COMM 'Correction values for device x'
//Define the transducer factor unit
CORR:TRAN:UNIT 'DB'
//Select the scale of the frequency axis
CORR:TRAN:SCAL LOG
//Define the data points of the transducer factor
CORR:TRAN:DATA 1MHZ,-10,10MHZ,-10,100MHZ,-15,1GHZ,-15
//Turn on the transducer
CORR:TRAN ON
//Automatically adjust the reference level
CORR:TRAN:ADJ:RLEV ON
Using Transducers